Reject Analysis of Extraoral Dental Radiography Examination at The Dental Hospital in Semarang, Indonesia

Marichatul Jannah(1*), Wingghayarie Patra Gandhi(2)


(1) Poltekkes Kemenkes Semarang, Indonesia
(2) Dental Hospital of Universitas Muhammadiyah Semarang, Indonesia.
(*) Corresponding Author

Abstract


Film rejection analysis is the process of calculating repeated exposure and determining the causes of repetition, so that they can be minimized or even eliminated. This study aims to determine the rejection ratio of films from extraoral dental radiography examinations, as well as the causes of each rejection percentage and efforts to reduce it. This study has a quantitative model with a descriptive approach. The research was conducted at the Radiodiagnostic Department of the Dental Hospital, Universitas Muhammadiyah Semarang in January-June 2022 with a total number of 232 radiographic film data. There were 4 radiographs that were rejected with the factors causing patient movement (50%), patient position (25%), and modality settings (25%). The film rejection ratio in the Radio-diagnostic Department of the Dental Hospital, Universitas Muhammadiyah Semarang is 1.7% (< 2%) so that a repair program is not needed.

Full Text:

PDF

References


Mallya S, Lam E. White and Pharoah’s Oral Radiology E-Book: Principles and Interpretation: Second South Asia Edition E-Book. Elsevier Health Sciences; 2019.

Russo P. Handbook of X-ray Imaging : Physics and Technology. CRC Press; 2017.

Kumar R, Khambete N, Priya E. Extraoral Periapical Radiography: An Alternative Approach to Intraoral Periapical Radiography. Imaging Sci Dent. 2011 Dec;41(4):165.

Papp J. Quality Management in The Imaging Sciences. 6th ed. Elsevier - Health Sciences Division; 2018.

Suraningsih N, Rosidah S, Felayani F. Analisa Penolakan Radiograf Di Instalasi Radiologi Rumah Sakit Bhayangkara Semarang. J Ilmu dan Teknol Kesehat. 2015;6(1):7–14.

Benza C, Damases-Kasi CN, Daniels ER, Amkongo M, Nabasenja C. The Causes of Reject Images in a Radiology Department at a State Hospital in Windhoek, Namibia. South African Radiogr. 2018 May 24;56(1).

Whaley JS, Pressman BD, Wilson JR, Bravo L, Sehnert WJ, Foos DH. Investigation of the Variability in the Assessment of Digital Chest X-ray Image Quality. J Digit Imaging. 2013 Apr;26(2):217–26.

Jabbari N, Zeinali A, Rahmatnezhad L, Jabbari N, Zeinali A, Rahmatnezhad L. Patient Dose from Radiographic Rejects/Repeats in Radiology Centers of Urmia University of Medical Sciences, Iran. Health (Irvine Calif). 2012 Feb 27;4(2):94–100.

Awad F, Naem F Al, Gemea A, Wedaa N, Mohammed Z, Elser ST. X-Ray Film Reject Analysis in Radiology Departments of Port Sudan Hospitals. Int J Radiol Imaging Tech. 2021 Jan 21;7(1).

Stephenson-Smith B, Neep MJ, Rowntree P. Digital Radiography Reject Analysis of Examinations with Multiple Rejects: an Australian Emergency Imaging Department Clinical Audit. J Med Radiat Sci. 2021 Sep 1;68(3):245–52.

Yurt A, Tintas M, Yüksel R. Reject Analysis in Digital Radiography: A Prospective Study. Int J Anat Radiol Surg. 2018 Oct 1;7(4):RO31–4.

IAEA. Radiation Protection in Dental Radiology, Safety Reports Series No. 108. Vienna; 2022.

Hofmann B, Rosanowsky TB, Jensen C, Wah KHC. Image Rejects in General Direct Digital Radiography. Acta Radiol Open. 2015 Oct 8;4(10).

Zennaro F, Oliveira Gomes JA, Casalino A, Lonardi M, Starc M, Paoletti P, et al. Digital Radiology to Improve the Quality of Care in Countries with Limited Resources: A Feasibility Study from Angola. PLoS One. 2013 Sep 25;8(9):e73939.

Kim TH, Heo DW, Jeong CW, Ryu JH, Jun HY, Han SJ, et al. Development of Portable Digital Radiography System with a Device for Monitoring X-ray Source-Detector Angle and Its Application in Chest Imaging. Sensors. 2017 Mar 7;17(3):531.

Jones & Bartlett Learning. Comprehensive Dental Assisting, Enhanced Edition. Jones & Bartlett Learning; 2020. 960 p.

Zewdu M, Kadir E, Berhane M. Analysis and Economic Implication of X-Ray Film Reject in Diagnostic Radiology Department of Jimma University Specialized Hospital, Southwest Ethiopia. Ethiop J Health Sci. 2017 Jun 29;27(4):421–6.

Alashban Y, Shubayr N, Alghamdi AA, Alghamdi SA, Boughattas S. An Assessment of Image Reject Rates for Digital Radiography in Saudi Arabia: A Cross-sectional Study. J Radiat Res Appl Sci. 2022 Mar 1;15(1):219–23.


Article Metrics

Abstract view : 204 times
PDF - 48 times

Refbacks

  • There are currently no refbacks.